Overview
HSIN-60Y Non-Invasive Sphygmomanometer Calibrator is a multi-functional test instrument used for oscillographic non-invasive blood pressure monitors. It provides ambulatory blood pressure simulation, static calibration, automatic leak testing, and overpressure release detection. This calibrator can be used to test various performance indicators of different blood pressure monitors. It simulates a pressure value of 450mmHg (60kPa) for automatic leak testing, overpressure testing, and pressure source simulation. It is suitable for measurement departments, hospitals, research institutes, university laboratories, etc., for the calibration of blood pressure gauges, precision pressure gauges, and general pressure gauges.
Feature
- Built-in air pump for high and low-pressure gas discharge inspection, leakage inspection, and pressure source generation.
- Automatic calculation and display of verification results for air tightness tests.
- Built-in standard diastolic and systolic blood pressure, heart rate can be automatically processed according to metrological verification procedures.
- Users can customize diastolic and systolic blood pressure, heart rate for measurement verification.
- Pulse intensity can be customized to simulate human pulse strength.
- Built-in or external cuff allows direct testing of non-invasive blood pressure monitors without a cuff.
- Imported pressure sensor and air pressure components ensure accurate zero-point adjustment.
- Built-in overpressure protection ensures safety and reliability.
- Quick selection of detection points and items with long-term storage settings.
- Color LCD display for enhanced visual appeal.
- Leak test function included.
- Pressure output function with pressure gauge detection capability.
- Ambulatory blood pressure simulation and customization functions allow users to set parameters including systolic and diastolic blood pressure, heart rate, etc.
Technical Parameter
◆ Static Pressure Measurement:
1. Pressure Range: (0-60) kPa[(0-450)mmHg]
2. Pressure Measurement Range: (0-60) kPa[(0-450)mmHg]
3. Accuracy Level: 0.1/0.05
4. Resolution: 0.01 kPa(0.1mmHg)
5. Pressure Control Stability: 0.05%F.S
◆ Dynamic Blood Pressure Simulation:
1. Adult Mode: Systolic Blood Pressure: (6.7 ~ 34.0)kPa[(50 ~ 255)mmHg],
Diastolic Pressure: (4.0 ~ 26.0)kPa[(30 ~ 195)mmHg]
2. Neonatal Mode: Systolic Blood Pressure: (4.0 ~ 16.0)kPa[(30 ~ 120)mmHg],
Diastolic Pressure: (1.3 ~ 13.3)kPa[(10 ~ 100) mmHg]
3. Repeatability of Blood Pressure: ≤0.3kPa(2mmHg)
4. Heart Rate Range: (30-250) times/min
5. Pulse Volume: 2.4cc at 200BPM, 1.2cc at 250BPM
6. Heart Rate Accuracy: ±(1%R.D+1) (30 ~ 200) times/min
7. Pulse Volume: (0.1-2.4)CC
8. Increment: 0.1CC
9. Display Unit: kPa, mmHg (user selectable)
10. Size: (280*216*125) mm
11. Weight: 6.5kg
Regulation
<JJG_692-2010 Non-Invasive Automatic Measurement Sphygmomanometer>