HSIN6800 Automatic pressure calibration system is a newly developed pressure sensor calibration and screening system by HSIN. The system adopts multiple sensor calibration technologies, and can calibrate multiple pressure sensors at one time.
1. System features: The HSIN6800 pressure sensor comprehensive screening system can screen 1-48 (expandable to 350) or more pressure sensors at one time according to customer needs2. Operation features: Humanized design of operation interface, fully realizing manual, automatic and free switching3. Software features: 1) Secondary development of software functions based on customer needs can fully meet each customer's requirements for software functions.2) It has the ability to automatically collect, process, store data, and print test reports in a specified format.4. Data storage: Test data is exported, saved, and printed through a USB serial port.
System composition | The system is composed of console, pressure control system, pressure connection port, software, data collector, computer, printer, etc. Note: the pressure control system adopts sensor pressure controller or Hsin series pressure control system. |
Measuringrange | (-0.1~21)MPa;(-0.1~40)MPa;(0~70)MPa; (the medium is air) (select the range according to the demand) |
Sensoraccuracy | 0.008%I.S; 0.01%I.S; 0.02%I.S; 0.01%F.S; 0.02%F.S; 0.05%F.S |
weight | About 80kg |
Test volume | 50 ~1,000 ccm |
Control | 0 ~100 % FS |
Pressure control volatility | <0.003%F.S |
Control speed | <15 second |
power supply | AC220V |
work environment | (-10~40 )℃;<90% RH |
Implementation procedures and specifications | ●JJG 860-2015 Verification regulation of pressure sensor (static);●JJG 882-2004 verification regulation of pressure transmitter;●JJG544-2011Verification regulation of pressure controller;●GJB4409A-2011General specification for pressure sensors●TSG R0003-2007Supervision Regulation on safety technology of simple pressure vessels●TSG ZF001-2006 Supervision Regulation on safety technology of safety valves●GB4208-84classification of shell protection grade●GB3096-2008Acoustic environment quality standard●GB/T12985General rule of applying percentile of human body size in product design●GB/T14775 General Ergonomic requirements for Manipulators●GJB-Z102Software reliability and safety design criteria |
Range selection table of 0.01% F.S pressure sensor | |||
measuring range | (-0.1~21)MPa | (-0.1~40)MPa | (0~70)MPa |
Gauge pressure | 0~2.5kPato 0~21MPa | 0~7kPato 0~40MPa | 0~10MPato 0~70MPa |
positive negativepressure | -1.25~1.25kPato -0.1~21MPa | -3.5~3.5kPato -0.1~40MPa | / |
absolutepressure | 0~50kPato 0~21.1MPa abs | 0~52kPato 0~40.1MPa | 0~10.1MPato 0~70.1MPa |
Range selection of 0.01% i.s-50 pressure sensor | |||
measuring range | (-0.1~21)MPa | (-0.1~40)MPa | (0~70)MPa |
Gauge pressure | 0~0.1MPato 0~21MPa | 0~0.1MPato 0~40MPa | 0~10MPato 0~40MPa |
positive negativepressure | -0.1~1MPato -0.1~21MPa | -0.1~1MPato -0.1~40MPa | / |
absolutepressure | 0~0.1MPato 0~21.1MPa abs | 0~0.1MPato 0~40.1MPa | 0~10.1MPato 0~40.1MPa |